We offer an assortment of refurbished and used Liquid Chromatography solutions ranging from HPLC, LC/MS to FPLC. Our systems can be customized to include almost any configuration of Autosamplers, Detectors and Pumps. Some of our leading manufacturers include Agilent, HP, Waters, GE, Amersham, Dionex and Shimazdzu. Our HPLC systems such as the Agilent 1100 and 1200 Series, Akta FPLCs, and the Waters Alliance & 600 series have the option of on site training, installation and extended warranties.

Uplc Binary Solvent Manager: 1-1 of 1

*Please Select a Category and/or Manufacturer dropdown OR Search By Model Name, # or Keyword.* Advanced Search >
Waters Acquity Binary Solvent Manager UPLC Binary Solvent Manager

Waters Acquity Binary Solvent Manager UPLC Binary Solvent Manager

Cambridge Scientific ID #'s: 15770, 15341


  • Number of solvents: Up to four: in any combination of two: Al or A2 and Bl or B2
  • Solvent storage: Solvents Tray accommodates up to four chromatographic solvents, two Sample Manager wash solvents and one Binary Solvent Manager seal wash solvent
  • Solvent conditioning: Vacuum degassing (Six-channel): one channel per solvent, and two channels for Sample Manager wash solvents
  • Operating flow rate range: 010 to 2.000 mL/min, in 0.001 mL increments
  • Compressibility compensation: Automatic and continuous
  • Effective system delay volume: <120 ulL, independent of system backpressure (with standard 50-ilL mixer installed)
  • Plunger seal wash: Integral, active, programmable
  • Gradient profiles: Eleven gradient curves [including linear, step (2), concave (4), and convex (4)]
  • Wet prime: Automatic
  • Maximum operating pressure: 15,000 psi up to 1 mL/min, 9,000 psi up to 2 mL/min per pump, not more than 15,000 psi total
  • Composition accuracy:
    • ±0.5% absolute (full scale) from 5 to 95% of flow rates
    • from 0.5 to 2.0 mLlmin (Contact Waters for conditions used)
  • Composition precision: 0.15% RSO or ±0.04 min SO, whichever is greater, based on retention time
  • Flow precision: 0.075% RSO or ±0.02 min SO, six replicates, based on retention time or volumetric measures (0.500 to 2.000 ml/min)
  • Flow accuracy: ± 1.0% at 0.5 mL/min with degassed methanol, per Waters AQT/SystemsQT protocol
  • Primary wetted materials: 316 stainless steel, UHMWPE, sapphire, ruby, FEP, PTFE, ETFE, diamond-like coating, PEEK and PEEK alloys, titanium alloys
  • Unattended operation: Leak sensors, full diagnostic data captured through console software
Learn More
By clicking below we can further assist you with your inquiry. Inquire 2 In Stock Worldwide Shipping