Specifications:
- Electrical
- Power Requirement: 100 to 130 VAC (60 Hz) or 200 to 250 VAC (50 Hz)
- Fuses (2): 250V time delay (Type T): 2-Amp (100-130 VAC), 1-Amp (200-250 VAC)
- Power Consumption: 150 watts
- Memory Backup: Integral lithium cell; 10 years typical
- Sample Volume: 0.20 to 0.25 mL
- Sample Capacity: Single sample
- Readout: 20-character digital display
- Units: mOsm/kg H2O
- Ranges:
- Low: 0 to 2000 mOsm/kg H2O
- High: 1400 to 4000 mOsm/kg H2O
- Resolution: 1 mOsm/kg H2O
- Communications: DTE RS-232C, parallel printer, barcode
- Performance at Reference Conditions
- Linearity: Less than ±0.5% from straight line
- Repeatability: 0 to 400 mOsm ±2 mOsm/kg (1 S.D.) 400 to 4000 mOsm ±0.5% (1 S.D.)
- Drift: Less than 1 digit per month
- Performance Over Reference Conditions
- Temperature Effect: 1 digit for every 5°C (9°F) ambient temperature change
- Operating Conditions Temperature:
- 18°C to 35°C (64°F to 95°F)
- Humidity: 5 to 80% R.H. (non-condensing)
- Storage Temperature: – 40°C to +45°C (-40°F to +113°F)
- Start-up Time: From Stand-By: immediately
- From Power On: under 5 minutes
- Test Time: Approximately 2 minutes per sample in 0 to 2000 mOsm range; moderately longer in high range.
- Dimensions
- 0 cm x 32.0 cm x 42 cm
- (11.5″ H x 12.5″ W x 16.5″ D)
- Height With Head Up: 41 cm (16″ H)
- Weight
- Net: 12 kg (27 lbs.)
- Shipping: 18 kg (38 lbs.)