Description
Specifications:
- Sample capacity: 100 x 2 ml vials in 1 tray. Microvials (100 or 300 µl) with sleeves (reduced cooling performance with microvials)
- Settable injection volume
- to 8 µl with small loop capillary
- to 40 µl with extended loop capillary
- Precision
- Typically < 0.5 % RSD from 5 to 40 µl
- Typically < 1 % RSD from 1 to 5 µl
- Typically < 3 % RSD from 0.2 to 1 µl
- Minimum sample volume: µl from 5 µl sample in 100 µl microvial, or 1 µl from 10 µl sample in 300 µl microvials
- Carryover
- Typically < 0.1 % without automated needle wash
- Typically < 0.05 % with external needle cleaning and 1 µl injection volume
- Sample viscosity range: 2 – 5 cp
- Recommended pH-range: – 8.5 solvent with pH < 2.3 should not contain acids which attack stainless steel. Upper pH range is limited by fused silica capillaries
- Material in contact with solvent: Stainless steel, sapphire, PTFE, PEEK, fused silica
- GLP features: Early maintenance feedback (EMF), electronic records of maintenance and errors
- Communications: Controller-area network (CAN). GPIB (IEEE-448), RS232C, APG-remote standard, optional four external contact closures and BCD vial number output
- Safety features: Leak detection and safe leak handling, low voltages in aintenance areas, error detection and display
- Housing: All material recyclable
Can be sold as an individual component or as part of a complete stack