Description
Includes Installation and One year Service Contract
Includes:
- G1946D LC/MSD Mass Spectrometer Series SL
- Roughing Pump
- Optional Front end HPLC
Specifications
- Mass Range: 2-3000m/z
- Mass Accuracy: +/-.13 u within the calibrated mass range in scan mode
- Mass Axis Stability: +- 0.13 u over 8 hours
- Scan Speed:
- 2500 u/s standard mode
- 5250 u/s fast scan mode
- SIM Sensitivity:
- ESI at 400 ul/min
- APCI at 1000 ul/min
- Selected ion monitoring of m/z 609.3
- Positive ionization
- Signal To Noise Ratio
- Quantity: 1 pg reserpine 100:1 RMS (20:1 peak-to-peak)
- Scan Sensitivity:
- ESI at 400 ul/min
- APCI at 1000 ul/min
- Scan range m/z 100-650
- Scan speed 2500 u/s
- Extracted ion at m/z 609.3
- Positive ionization
- Signal To Noise Ratio
- Quantity: 50 pg reserpine 100:1 RMS (20:1 peak-to-peak)
- Multiple Signal Acquisition: Ability to cycle through four different acquisition modes on a scan-by-scan basis within a single run