Description
Features:
- G1364C Fraction Collector, analytical scale, designed for flow rates below 10 ml / min. (can be modified for higher flow rates) and for the use with vials, test tubes of up to 75 mm height, well-plates and a 40-funnel tray connecting to external locations of any size
- For analytical scale applications
Specifications:
- Trigger Modes: Time slices, Peak (threshold, up- /downslope), Timetable (combination of time intervals and peak) and Manual trigger (supported only with G1323BControl Module)
- Operating Modes: Discrete fractions: default mode for all vessels. The flow is diverted to waste, while moving from one vessel position to the next vessel position Continuous flow: optional, available only when using well plates. It is possible to move from one well plate position to the next one without diverting the flow into the well plate to waste
- Test tube /plate sizes: Minimum 48 mm to 100 mm maximum
- Maximum tube volume: ca. 45 ml
- Maximum flow rate: 100 ml / min (depending on viscosity and generated back pressure, max. 6 bar at the diverter valve)
- Delay volumes [μl]:
- Fraction collector inlet to diverter valve: -500 (typical, depends on length of the tubing)
- Diverter valve: ~15
- Diverter valve to needle: ~110
- Needle: -~5
- Delay calibration Sensor: Single wavelength absorbance detector working at 654 nm, consisting of a LED and a photo diode
- Diverter Valve: 3/2 Diverter valve with low internal volume (15 μl), switching time < 100 ms, maximum operating pressure 6 bar
- Cooling: Optional (with additional G1330B), performance depending on ambient conditions and the volume of collected fractions
- Maximum capacity: 3 fraction collectors in parallel plus one recovery fraction collector connected via 12-Position, 13-Port Selector valve
- GLP features Early maintenance feedback (EMF), electronic records of maintenance and errors
- Interfaces: Controller-area network (CAN). optional; LAN or external contacts interface RS232C, APG-remote (for remote start / stop signals to / from other modules) Interface to G1330A Thermostat CAN-DC-out for operation of Agilent approved external devices like valves
- Safety features: Leak detection and safe leak handling, error detection and display, exhaust fan for fume extraction of hazardous vapors
Can be sold as an individual component or as part of a complete stack