Description
Features:
The detector is designed for highest optical performance, GLP compliance and easy maintenance.
Specifications:
- Detection Type: 1024-element photodiode array
- Light Source: Deuterium and tungsten lamps The UV-lamp is equipped with RFID tag that holds lamp typical information.
- Data Rate: up to 80 Hz
- Wavelength Range: 190 – 950 nm
- Short Term: noise (ASTM) Single and Multi-Wavelength: < ± 0.7·10-5 AU at 254 and 750 nm
- Drift: < 0.9·10-3 AU/h at 254 nm
- Linear Absorbance Range: > 2 AU (5 %) at 265 nm
- Wavelength Accuracy: ± 1 nm ,Self-calibration with deuterium lines, verification with holmium oxide filter
- Wavelength Bunching: 1 – 400 nm, Programmable in steps of 1 nm
- Slit Width: 1, 2, 4 , 8, 16 nm Programmable slit
- Diode Width: < 1 nm