Agilent 1260 Series – G7114A HPLC Variable Wavelength Detector

Cambridge Scientific ID: 28723 Category: Tag: Brands:

Description

Product Description G7114A

The Agilent 1260 Infinity II Variable Wavelength Detector (VWD) is the most sensitive and fastest detector in its class. Time-programmable wavelength switching provides sensitivity and selectivity for your applications. More sample information can be acquired in the dual wavelength mode. Low detector noise (<±1.5 µAU) and baseline drift (< 1 x 10-4 AU/h) facilitates precise quantification of trace levels components. High productivity can be achieved with fast analysis at up to 120 Hz data rates.

Specifications:

Physical Specifications;

  • Weight:  11 kg (24.3 lbs)
  • Dimensions (height × width × depth): 140 x 396 x 436 mm (5.5 x 15.6 x 17.2 inches)
  • Line voltage:  100 – 240 V~, ± 10 %
  • Line frequency: 50 or 60 Hz, ± 5 %
  • Power consumption: 80 VA, 70 W
  • Ambient operating temperature: 4 – 55 °C (39 – 131 °F)
  • Ambient non-operating temperature: -40 – 70 °C (-40 – 158 °F)
  • Humidity: < 95 % r.h. at 40 °C (104 °F)
  • Operating altitude:  Up to 3000 m (9842 ft)
  • Non-operating altitude: Up to 4600 m  (15091 ft)

Performance Specifications;

  • Detection type: Double-beam photometer
  • Light source: Deuterium lamp
  • Number of signals: Single and dual wavelength detection
  • Maximum date rate: 
      • 120 Hz (single wavelength detection)
      • 2.5 Hz (dual wavelength detection)
  • Short term signal noise (ASTM): 
      • <±0.25·10-5 AU, at 230 nm (single wavelength detection)
      • <±0.80·10-5 AU, at 230 nm and 254 nm (dual wavelength detection)
  • Drift: <1·10-4 AU/h, at 230 nm
  • Linear absorbance range: >2.5 AU upper limit
  • Wavelength range: 190 – 600 nm
  • Wavelength accuracy: ±1 nm, self-calibration with deuterium lines, verification with holmium oxide filter
  • Wavelength precision: <±0.1 nm
  • Slit width: 6.5 nm typical over whole wavelength range
  • Time programmable: Wavelength polarity, peak width, lamp on/off
  • Spectral tools: Stop-flow wavelength scan
  • Analog output: Recorder/Integrator 100 mV or 1 V, 1 output
  • Instrument Control
      • Lab Advisor B.02.08 or above
      • LC and CE Drivers A.02.14 or above
      • For details about supported software versions refer to the compatibility matrix of your version of the LC and CE Drivers
  • Communication: 
    • LAN, Controller Area Network (CAN), USB
    • Extended Remote Interface: ready, start, stop and shut-down signals

Contact Us

Thank you for your interest in the Agilent 1260 Series – G7114A HPLC Variable Wavelength Detector (SKU:28723) To speak with a representative, please fill out the contact form below or call us directly.

Product Inquiry