For more in-depth information from every sample scanned, we gave the Odyssey M a sophisticated sCMOS image sensor and industry-leading 5-micron resolution. No other imager can give you this level of sensitivity and dynamic range. With NIR and visible fluorescence, RGB, and optional luminescence capabilities—all 19 channels calibrated and integrated for consistent performance—you can easily image an ever-expanding variety of assays. And our innovative line-scanning technology leads the industry in speed. The Odyssey M acquires all this data in minutes, not hours.

Specifications: 

  • Laser:
      • Laser Source: 488 nm, 520 nm, 685 nm, 785 nm.
      • Maximum Laser Output: 30 mW, 50 mW, 450 mW, 1.2 W
      • Beam Divergence; 0.038°, 0.046°, 0.085°, 0.060°
  • Strong Electromagnetic Fields: Operating the Odyssey M in the presence of a strong electromagnetic field in the 100MHz to 400MHz frequency range has been shown to degrade the performance of the chemiluminescent imager. It is possible that some allowable faint bars or striping may be seen at slightly more than the background levels of the camera. When used in the expected environment, this interference should be uncommon.
  • LED: LED Risk Group: Exempt group in accordance with IEC 62471:2006. The LEDs do not pose any photobiological hazard.
  • Focus: Focus offset range: -1.00 mm to 5.00 mm
  • Pixel Resolution:
      • Available imaging resolutions: 5 μm, 10 μm, 20 μm, 50 μm, 100 μm
      • Resolution for chemi images: 100 μm
  • Maximum Height of Samples: To image an item on the Odyssey M, the item must be 23 mm or less from the Scan Surface to the top of the item.
  • Operating Conditions: For indoor use only; operating temperature 15 – 30 °C and dew point not greater than 22 °C, non-condensing; altitude not to exceed 2000 m.
  • Transport Conditions: Shipping/transport conditions for the Odyssey M Imager should not exceed -40 °C to 65 °C.
  • Environmental Conditions: Pollution Degree 2
  • Power Requirements: 
    • Universal input range is between 100-240 VAC (voltage fluctuations not to exceed 10% of the nominal voltage); < 3 Amp maximum; At 120 V: 0.5 Amp typical when idle; 50-60 Hz.
    • Power Supply: One end of the provided power supply connects to a wall outlet and the other end connects to the imager. The power supply converts from wall outlet voltage to 24 VDC (5 Amp). Only use the provided power supply to plug in the imager. If you need another power supply, contact LICORbio.
  • Dimensions: 
      • 24″ width × 30″ depth × 15″ height (61 cm width × 76 cm depth × 38 cm height)
      • Height with lid fully open: 28″ (71 cm)
  • Scan Surface: 
      • Total Scan Area: 25 cm W × D 18 cm (9.8” W × 7.1” D)
      • Scan Area of Chemi Region (Odyssey M Model 3350 only): 15 cm W x 11 D cm (5.9” W × 4.3” D)
  • Dynamic Range: >6 logs for chemiluminescence (Model 3350) and fluorescence. This dynamic range is obtained in a single acquisition. 
  • Weight:
      • Odyssey M Model 3350: 121 lbs (55 kg)
      • Odyssey M Model 3340: 115 lbs (52 kg)
  • Network: 
      • Protocol: TCP/IP
      • Connection: Use the Category 6 network cable provided with the Odyssey M.
  • Detectors: 
      • sCMOS image sensor
      • Sensor for chemiluminescence (Model 3350): CCD
  • Expected Laser Lifetime: 
      • 685 nm: 20,000 hours
      • 785 nm: 20,000 hours
      • 488 nm: 40,000 hours
      • 520 nm: 40,000 hours
  • Light Sources:
    • RGB LED (trans-illumination)
    • RGB LED (reflective illumination)
    • Solid-state diode laser at 488 nm
    • Solid-state diode laser at 520 nm
    • Solid-state diode laser at 685 nm
    • Solid-state diode laser at 785 nm

Contact Us