Memmert ICH260 Environmental Chamber Incubator

SKU: 34923 Category: Tag: Brands:

Description

ICH260 Memmert stability climate chambers guarantee the perfect atmosphere for climate and temperature tests. Our constant climate chamber is tailored to environmental simulation, material testing and stability testing in accordance with ICH guidelines. Former Models: ICH260, ICH600.

Features:

  • Working temperature range without humidity: −10 °C up to 60 °C
  • Working temperature range with humidity: 10 °C up to 60 °C
  • Setting accuracy temperature: 0.1 °C
  • Temperature: 2 Pt100 sensors Class A in 4-wire-circuit, mutually monitoring and taking over the performance at the same temperature value
  • Humidity supply with distilled water from external tank by self-priming pump
  • Humidification by hot steam generator
  • Active humidifying and dehumidifying adjustable from 10–80 % rh, with digital display of relative humidity (resolution 0.1 %, setting accuracy 0.5 %)
  • ControlCOCKPIT: TwinDISPLAY, adaptive multifunctional digital PID-microprocessor controller with 2 high-definition TFT-color displays
  • Language settings: German, English, Spanish, French, Polish, Czech, Hungarian
  • Adjustable parameters: temperature (°C or °F), relative humidity, program time, time zones, daylight savings time
  • Timer: digital backwards counter with target time setting, adjustable from 1 minute to 99 days
  • Function SetpointWAIT: process time does not start until the set temperature is reached
  • Calibration: three freely selectable values each, temperature / humidity / CO₂

Safety:

  • Temperature control: mechanical temperature limiter TB, protection class 1 according to DIN 12880, switches off heating approx. 10 °C above nominal temperature
  • Temperature control: over- and under-temperature monitor TWW, protection class 3.3, or adjustable temperature limiter TWB, protection class 2, adjustable at the controller
  • AutoSAFETY: integrated over- and under-temperature protection “ASF”, automatically follows setpoint value within preset tolerance range; alarms for over/under temperature; heating switches off in case of overtemperature, compressor in case of undertemperature
  • Autodiagnostic system: integral fault diagnostics for temperature and humidity control
  • Alarm: visual and acoustic alarms for over-/under-temperature, under humidity, open door, and empty water tank

Electrical Data:

  • 230 Volts
  • 50 Hz / approx. 1350 W

Interior & Exterior:

  • Stainless steel interior
  • Volume: 256 L
  • Dimensions W × H × D (mm): 640 × 800 × 500
  • Max. number of internals: 9
  • Max. loading of chamber: 200 kg
  • Max. loading per internal: 20 kg
  • Textured stainless steel casing
  • Dimensions W × H × D (mm): 824 × 1552 × 734
  • Housing: rear zinc-plated steel
  • Installation: on lockable castors

Stainless Steel:

For many years, exterior textured stainless steel has been the unmistakable feature of Memmert ovens, incubators, and climate chambers in laboratory and light-industrial settings. Memmert equipment exemplifies functional design in its most elegant and hygienic form, while still being scratch-resistant, robust, and durable.

Stainless steel is the ideal material for temperature control chambers and heating baths. It is particularly corrosion-resistant, hygienic, robust and recyclable without question.

For the outer housing of all Memmert appliances, textured stainless steel is used (rear walls are zinc-plated sheets). The inner chamber, sliding shelves/grids, housing, interior basin and accessories for every heating bath use stainless steel of class W-St. 1.4301 (ASTM 304). Since aggressive substances are frequently used in the vacuum drying oven, the inner chamber and thermoshelves are made of stainless steel W-St. 1.4404 (ASTM 316L), and the tubing is made of stainless steel W-St. 1.4571 (ASTM 316Ti).