QIAcube Connect offers automation for over 80 QIAGEN kits with up to 12 samples per run. It was developed by five teams with members from Germany, Switzerland, UK, India and the USA. With it, QIAGEN now offers a breakthrough system with advanced digital capabilities and connectivity. It builds on the success of ~3000 QIAcube instruments that already allow research labs to fully automate and track their well-known QIAGEN spin-column protocols with just a push of a button. 

  • Technical data: Power: 100–240 V AC, 50/60 Hz, 650 VA Mains supply voltage fluctuations are not to exceed 10% of nominal supply voltages. Note: The apparent power can exceed 650 VA for up to 2 seconds during the centrifuge acceleration and can reach an approximate value of 1000 VA.; Fuse: 2x T6.3A L 250V; Overvoltage category: II
  • Operating temperature: 18 to 28ºC (64.4 to 82.4ºF)
  • Relative humidity: 15–75% (non-condensing)
  • Place of operation: For indoor use only
  • Pollution level: 2
  • Environmental class: 3K2 (IEC 60721-3-3)
  • Storage conditions: Temperature: 15ºC to 30ºC (59ºF to 86ºF) in manufacturer’s package; Relative humidity: Max. 75% (non-condensing); Environmental class: 1K2 (IEC 60721-3-1)
  • Instrument dimensions: Dimensions (hood closed): width 65 cm (25.6 in.), height 58 cm (22.8 in.), depth 62 cm (24.4 in.); Dimensions (hood open): width 65 cm (25.6 in.), height 86 cm (34 in.), depth 62 cm (24.4 in.); Mass: QIAcube Connect: 71.5 kg (157.6 lb.), Accessories: 3 kg (6.6 lb.)
  • Centrifuge: 10,640 rpm maximum, 12,000 x g maximum, swing-out rotor maximum 45°, 12 rotor positions
  • Shaker: Speed 100–2000 rpm, amplitude 2 mm, heating range of ambient temperature to 70°C (158°F), ramp-up time of <5 minutes from ambient temperature to 55°C (±3°C). Difference in the temperature detected by the internal sensor and the temperature of the sample liquid is approximately –2°C.
  • Pipetting system: Syringe size 1 ml, pipetting range 5–900 µl
  • Table capacity: Up to 12 samples per run
  • Touchscreen: 10.4” TFT Touchscreen, active area 211.2 x 158.4 mm, resolution 800*600 SVGA
  • Data output: USB stick: USB2.0; Compatible OS: Windows 8, Windows 7, Windows Vista, Windows XP (SP3 or later); Mac OS x 10,1 or later; Operating temperature range: 0ºC to 35ºC; Operating humidity range: 10 to 90% (with no condensation); Storage temperature range: –20ºC to 60ºC (–4ºF to 140ºF); Storage humidity range: 10 to 90% (with no condensation); Formatting: FAT32
  • UV lamp: Wavelength: 278nm; Optical power: 200-300mW
  • Special feature: Handheld scanner; Scan pattern: Area Image (838 x 640-pixel array); Motion tolerance: Up to 610 cm/s (240 in/s) for 13 mil UPC at optimal focus; Symbol contrast: 20% minimum reflectance difference; Decode capability: Reads standard 1D, PDF, 2D, Postal and OCR symbology
  • Software: QIAGEN protocols are pre-installed on the QIAcube Connect.
  • Altitude: Up to 2000 m (6500 ft)
  • Transportation conditions: –25°C to 60°C (–13°F to 140°F) in manufacturer’s package; Max. 75% relative humidity (non-condensing); Environmental class: 2K2 (IEC 60721-3-2)