Description
To be used with a Shimadzu 10 HPLC System
Specifications:
- Sample injection system: Total sample injection, variable sample injection system (zero sample loss during injection)
- Settable range of injection volumes: to 100 µL (standard), or 1 to 2000 µL (option) (0.1µL steps from 0.1 to 0.9µL, and 1µL steps from 1 to 2000 µL)
- Sample processing capability:
- With 1-mL vials: 350
- With 1.5-mL vials: 140
- with 4-mL vials:100
- With micro plates: 4 (96 wells: 284, 284 wells: 1536)
- with deep well plates: 4 (96 wells: 284, 284 wells: 1536)
- Injection volume repeatability: RSD: 0.3% max. (at 10µL injection)
- Cross contamination: 01% max. (for analysis of naphthalene)
- Injection volume accuracy:Less than +/- 1% (50µL)
- Injection volume accuracy correction: Possible
- of injections: 1-99 times per sample
- Setting of analysis time: .01– 9999.9 (0.1 minute increments to 999.99 min., 0.1 minutes increments over 1000 min.)
- Number of steps for injection conditions: Max 102 steps
- Sample suction speed: 1-15 µL/sec (0.1µL/s step for 0.1-0.9 µL/sec., 1µL/s step for over1 µL/sec.)
- Rinsing liquid suction speed: 1-35 µL/sec/ (1µL/s step)
- Maximum pressure: 20.0 Mpa (standard pippings of sample flowpath is PEEK)
Sample cooler (SIL-HTc only)
- Method: Direct cooling method (at a setting of 4 deg. C, room temperature must be less than 30 deg. Cand humidity less than 70%)
- Temperature setting range: 4 to 40 deg. (cooling down to 4 deg. C for room temperature up to 30 deg. C and humidity to 70%)
- Operating ambient temperature range:4 to 35 deg. C