Description
The EC10 is an advanced temperature testing chamber intended for automated test system and laboratory applications. The .7 ft3chamber uses vertical air flow with convection style heating and has a temperature range of -184°C to 315°C (using LN2) and -73°C to 315°C (using LCO2). The EC10 model temperature chamber uses the advanced and extremely flexible TC10 controller to provide features such as non-volatile memory to store user temperature profiles and parameters, an additional probe (for monitoring and control of product temperature), adjustable ramp rate, power down restart, analog and auxiliary I/O. Built in IEEE-488 (GPIB), RS232/RS422 remote interfaces are standard features on the EC10 chamber. The chamber can function on it’s own in an Automated Test Environment. Standalone operation is supported by a full function keyboard with a 2 line, 16 character LCD display. The electrically isolated chamber temperature and user temperature probe readings are normally displayed continuously and either probe may be used for control to ensure the proper temperature is achieved for the DUT. User programs are easily entered into the EC10 chamber using a BASIC like command set programming language and stored in non-volatile memory along with user adjustable controller parameters. Probe calibration procedures are built in using a local menu driven format and parameters such as the PID coefficients are adjustable from the keyboard and remote interfaces. In addition, alarm function and sound level, BAUD rate, interrupt assignments and other communication port options are configurable from the front panel using an easy, menu driven format.
Specifications:
- Power: 1600 WATTS
- Voltage (frequency): 110/220 VAC (50/60Hz)
- Heat/Cool Rate: 30C/min
- Air Flow: 60 CFM
- Workspace Dimensions: 12″(W) x 9.75″(H) x 10.25″(D)
- Overall Dimensions: 21″(W) x 16″(H) x 24″(D)
- Volume: 0.7 cubic feet
- Weight: 49 lbs.