Description
Features:
- Optimized for high-throughput applications and sub two micron column particles
- Maximum reproducibility with isothermal injection and separation
- Greater application flexibility with multiple sample formats
- Superior robustness with a diamond coated high pressure valve Accela Photo Diode Array Detector
- Highest detection sensitivity with proprietary LightPipe technology
- Designed for U-HPLC separations with a 1cm path length, 2µL LightPipe flowcell
- Maximum peak resolution with fiberoptic beam shaper
- Easy maintenance with integrated wavelength validation
Specifications:
- Sample capacity:
- 200 vials in five removable trays
- 288 samples in three 96-well microplates
- 1152 samples in three 384-well microplates
- Standard vial:8 mL volume
- Injection precision: < 1.0% RSD at 5 μL and up
- Partial loop injection volume:
- 10 μL with standard 25 μL sample loop and 250 μL syringe
- Up to 1000 μL injection volume with optional loops and syringes
- Fixed loop (full loop) injection volume:
- 25 μL with standard loop and 250 μL syringe
- Up to 1000 μL injection volume with optional loops and syringes
- Sample carryover:
- < 0.1%
- 6 mL flush volumes, user-specified for internal and external needle wash cycles
- Minimum sample volume: 1 μL can be injected from 5 μL
- Needle height control:
- Programmable in 0.1 mm increments
- Active vial bottom search selectable on/off
- Column oven: Programmable in 1 °C increments from 5 to 95 °C
- Tray temperature control:
- Programmable in 1 °C increments from 0 to 60 °C
- Typical vial temperature from 4 to 50 °C
- Reagent vials: Four 16 mL vials can be specified for flush/wash or dilution functions
- Remote inputs: Pump ready, inject hold
- Remote outputs: Pump stop, autosampler ready, inject, gradient start, four timed events
- Ambient environment:10 to 40 °C, 5 to 95% relative humidity, non-condensing
- Power requirements: 100/120/220/240 Vac nominal; 50/60 Hz, 550 VA