Used Lab Equipment

Used Lab Equipment

At Cambridge Scientific Products, we offer an extensive range of equipment for your laboratory. Interact with our search tool below to explore our current products, or filter by a product category to see our most up to date product offerings in each of these categories.

Agilent 1100 Series - G1367A HPLC Well Plate Autosampler

*Please Select a Category and/or Manufacturer dropdown OR Search By Model Name, # or Keyword.* Advanced Search >
Agilent 1100 Series - G1367A HPLC Well Plate Autosampler

Agilent 1100 Series - G1367A HPLC Well Plate Autosampler

Cambridge Scientific ID #: 22334

Specifications:

  • GLP features: Early maintenance feedback (EMF), electronic records of maintenance and errors
  • Communications Controller: area network (CAN). RS232C, APG-remote standard, optional four external contact closures and BCD vial number output
  • Safety features:  Leak detection and safe leak handling, low voltages in maintenance areas, error detection and display
  • Injection range:  
    • 1 – 100 μl in 0.1 μl increments
    • Up to 1500 μl with multiple draw (hardware modification required)
  • Precision:
    • Typically < 0.5 % RSD of peak areas from 5 – 100 μl,
    • Typically < 1 % RSD from 1 – 5 μl
  • Sample viscosity range:  0.2 – 5 cp
  • Sample capacity:
    • 2 × well-plates (MTP) + 10 × 2 ml vials
    • 100x2ml in one tray
    • 40 x 2 ml in half tray
  • Injection cycle time:
    • Typically < 30 s using following standard conditions:
    • Default draw speed: 200 μl/min
    • Default eject speed: 200 μl/min
    • Injection volume: 5 μl
  • Carry-over Typically:
    • < 0.01 % using the following conditions:
    • Column: 125 x 4 mm Hypersil ODS, 5 μm
    • Mobile phase: Water/Acetonitrile = 80/20
    • Flow rate: 1 ml/min
    • Injection volume: 1 μl caffeine (1 mg/ml), 5 μl water to test carryover
    • Outside wash of needle before injection: 20 sec with water using flush port

 

In Stock Worldwide Shipping Call: 1-877-Used-Lab

Our used lab equipment is fully refurbished, guaranteed to work up to manufacturers specifications and are delivered to our customers in as close to new condition as possible. Typical turnaround for refurbishing of used lab equipment is 2-4 weeks. They also come with our standard 6 Month parts and labor warranty. Prices do not include S/H unless otherwise indicated. Pictures enclosed and the actual unit may vary unless specified.

Inquire

Contact Us:

Thank you for your interest in the Agilent 1100 Series - G1367A HPLC Well Plate Autosampler (ID #: 22334). To speak with a representative, please fill out the contact form below or call us directly.