Agilent 1100 Series - G1377A HPLC Well Plate Autosampler

*Please Select a Category and/or Manufacturer dropdown OR Search By Model Name, # or Keyword.* Advanced Search >
Agilent 1100 Series - G1377A HPLC Well Plate Autosampler

Agilent 1100 Series - G1377A HPLC Well Plate Autosampler

Cambridge Scientific ID #: 13766

Specifications:

  • GLP features: Early maintenance feedback (EMF), electronic records of maintenance and errors
  • Communications: Communications Controller-area network (CAN). RS232C, APG-remote standard, optional four external contact closures and BCD vial number output
  • Safety features: Leak detection and safe leak handling, low voltages in maintenance areas, error detection and display
  • Injection range:
    • – 8 µl in 0.01 µl increments with the small loop capillary
    • – 40 µl in 0.01 µl increments with the extended loop capillary
  • Precision:
    • Typically < 0.5 % RSD of peak areas from 5 – 40 µl,
    • Typically < 1 % RSD from 1 – 5 µl
    • Typically < 3 % RSD from 0.2 – 1 µl
  • Sample viscosity range: 2 – 5 cp
  • Sample capacity:
    • 2× well-plates (MTP) + 10 × 2 ml vials
    • 100 x 2 ml in one tray
    • 40 x 2 ml in half tray
  • Injection cycle time:
    • Typically, < 30 s using following standard conditions:
    • Default draw speed: 4 µl/min
    • Default eject speed: 10 µl/min
    • Injection volume: 0.1 µl
  • Carry-over:
    • Typically < 0.05 % using the following conditions:
    • Column: 150 x 0.5 mm Hypersil ODS, 3 µm
    • Mobile phase: Water/Acetonitrile = 85/15
    • Column Flow rate: 13 µl/min
    • Injection volume: 1 µl caffeine (=25ng caffeine), 1 µl water to test carryover
    • Outside wash of needle before injection: 20 sec with water using flush port
In Stock Worldwide Shipping Call: 1-877-Used-Lab

Our used lab equipment is fully refurbished, guaranteed to work up to manufacturers specifications and are delivered to our customers in as close to new condition as possible. Typical turnaround for refurbishing of used lab equipment is 2-4 weeks. They also come with our standard 6 Month parts and labor warranty. Prices do not include S/H unless otherwise indicated. Pictures enclosed and the actual unit may vary unless specified.