Used Lab Equipment

Used Lab Equipment

At Cambridge Scientific Products, we offer an extensive range of equipment for your laboratory. Interact with our search tool below to explore our current products, or filter by a product category to see our most up to date product offerings in each of these categories.

Dionex Ultimate 3000 TCC-3200 HPLC Column Compartment

*Please Select a Category and/or Manufacturer dropdown OR Search By Model Name, # or Keyword.* Advanced Search >
Dionex Ultimate 3000 TCC-3200 HPLC Column Compartment

Dionex Ultimate 3000 TCC-3200 HPLC Column Compartment

Cambridge Scientific ID #: 18418

Specification:

 

  • Temperature Range: 5 °C to 85 °C (max. 18 °C below ambient)
  • Temperature Accuracy: ±0.5 °C
  • Temperature Stability: ±0.1 °C
  • Temperature Precision: ±0.1 °C
  • Column Capacity: Up to 6 columns, max. 30 cm length
  • Heat-up Time: Typically 15 minutes 20 °C to 50 °C
  • Cooldown Time: Typically 18 minutes 50 °C to 20 °C
  • Control: All functions software controlled via USB
  • I/O interfaces: 2 digital inputs, 2 programmable relay outputs
  • Switching valves: 1 or 2 thermostatted valves (2-position, 6-port, 2-position, 10-port, or 6-position, 7-port)

 

  • Watted parts

 

    • Stainless steel (1.4571, 316L), PEEK; Waste line: PTFE
    • Column switching valve (stainless steel): Vespel and DuraLife
    • Column switching valve (biocompatible): PEEK
    • Column switching valve (6-position, 7-port): Titanium
  • Safety features: Humidity sensor and gas sensor. Temperature control and temperature sensors

 

  • User input/display

 

    • LCD-indicating system parameters
    • Standby button
    • 3 LEDs for status monitoring (Power, Connected, Status)
    • 4 function keys for operation during initial installation and maintenance
  • GLP features: Electronic identification system for 4 columns. In Chameleon: Full support of automatic equipment qualification (AutoQ) and System System Wellness monitoring. All system parameters are recorded in the Chromeleon Audit Trail.
  • Power requirements: 100 - 120 V, 60 Hz; 200 - 240 V, 50 Hz, max. 150 VA
  • Environmental conditions: Indoor use only; 10 to 35°C (50 to 95°F) operating temperature range; 80% relative humidity (non-condensing); Overvoltage category 2; Pollution degree 2.
  • Dimensions (H x W x D): 17 x 42 x 51 cm (6.7 x 16.5x 20 in)  
  • Weight: 13.2 kg (29.0 lbs)
In Stock Worldwide Shipping Call: 1-877-Used-Lab

Our used lab equipment is fully refurbished, guaranteed to work up to manufacturers specifications and are delivered to our customers in as close to new condition as possible. Typical turnaround for refurbishing of used lab equipment is 2-4 weeks. They also come with our standard 6 Month parts and labor warranty. Prices do not include S/H unless otherwise indicated. Pictures enclosed and the actual unit may vary unless specified.

Inquire

Contact Us:

Thank you for your interest in the Dionex Ultimate 3000 TCC-3200 HPLC Column Compartment (ID #: 18418). To speak with a representative, please fill out the contact form below or call us directly.