At Cambridge Scientific Products, we offer an extensive range of equipment for your laboratory. Interact with our search tool below to explore our current products, or filter by a product category to see our most up to date product offerings in each of these categories.
Features:
The YSI 2300 STAT PLUS Glucose and Lactate Analyzer is a laboratory instrument intended for use in clinical care and sports medicine applications. It provides quick measurements of glucose in whole blood, plasma or serum; and of L-lactate in whole blood, plasma, or cerebrospinal fluid (CSF). In whole blood or plasma, glucose and L-lactate can be measured simultaneously.
Specifications:
Normal Mode: Displayed/printed result in 65 seconds
Sample to sample interval is 100 seconds
Screen Mode: Displayed/printed result in 45 seconds
Sample to sample interval is 70 seconds
Normal Mode:
Glucose: to 50.0 mmol/L (900 mg/dL or 9000 mg/L)
Lactate: to 30.0 mmol/L (267 mg/dL or 2670 mg/L)
Screen Mode:
Glucose: to 27.8 mmol/L (500 mg/dL or 5000 mg/L)
Lactate: to 15.0 mmol/L (134 mg/dL or 1335 mg/L)
Glucose: 10.0 mmol/L; 180 mg/dL; 1800 mg/L
Lactate: 5.00 mmol/L; 45 mg/dL; 445 mg/L
Glucose: 21 days (YSI 2365)
Lactate: 14 days (YSI 2329)
Glucose: ±2% of the reading or 2.5 mg/dL (25 mg/L, 0.2 mmol/L),whichever is larger
Lactate: ±2% of the reading or 0.1 mmol/L (1 mg/dL, 10 mg/L), whichever is larger
Glucose: 1 mg/dL (1 mg/L, 0.1 mmol/L)
Lactate: 0.1 mmol/L (1 mg/L, 1 mg/dL)
Our used lab equipment is fully refurbished, guaranteed to work up to manufacturers specifications and are delivered to our customers in as close to new condition as possible. Typical turnaround for refurbishing of used lab equipment is 2-4 weeks. They also come with our standard 6 Month parts and labor warranty. Prices do not include S/H unless otherwise indicated. Pictures enclosed and the actual unit may vary unless specified.