Description
Specifications:
- GLP features: Early maintenance feedback (EMF), electronic records of maintenance and errors
- Communications: Communications Controller-area network (CAN). RS232C, APG-remote standard, optional four external contact closures and BCD vial number output
- Safety features: Leak detection and safe leak handling, low voltages in maintenance areas, error detection and display
- Injection range:
- – 8 µl in 0.01 µl increments with the small loop capillary
- – 40 µl in 0.01 µl increments with the extended loop capillary
- Precision:
- Typically < 0.5 % RSD of peak areas from 5 – 40 µl,
- Typically < 1 % RSD from 1 – 5 µl
- Typically < 3 % RSD from 0.2 – 1 µl
- Sample viscosity range: 2 – 5 cp
- Sample capacity:
- 2× well-plates (MTP) + 10 × 2 ml vials
- 100 x 2 ml in one tray
- 40 x 2 ml in half tray
- Injection cycle time:
- Typically, < 30 s using following standard conditions:
- Default draw speed: 4 µl/min
- Default eject speed: 10 µl/min
- Injection volume: 0.1 µl
- Carry-over:
- Typically < 0.05 % using the following conditions:
- Column: 150 x 0.5 mm Hypersil ODS, 3 µm
- Mobile phase: Water/Acetonitrile = 85/15
- Column Flow rate: 13 µl/min
- Injection volume: 1 µl caffeine (=25ng caffeine), 1 µl water to test carryover
- Outside wash of needle before injection: 20 sec with water using flush port