Description
Features:
The Advanced Instruments Model 3320 Micro Osmometer determines the osmolality of solutions using freezing point depression (FPD). It is Ideal for serum, urine or other bodily fluids in clinical applications; and media or formulations in pharmaceutical, biotech or industrial applications. The instrument is automated, single sample designed and processes a 20 ?L sample within a one-minute test time. This allows the Model 3320 to provide the best accuracy and precision in the market today.
Specifications:
- Linearity: Less than ±1% from a straight line over calibrated range
- Repeatability: 0 to 400 mOsm: ±2 mOsm/kg (1 S.D.) 400 to 2000 mOsm: ±0.5% (1 S.D.)
- Drift: Less than 1 mOsm/kg H2O per month
- Temperature Effects: Less than 1 mOsm/kg H2O for every 5°C (9°F) ambient temperature change
- Operating Conditions Temperature: 18°C to 35°C (64°F to 95°F)
- Humidity: 5 to 80% relative humidity (non-condensing)
- Storage Temp.: -40°C to +45°C (-40°F to +113°F)
- Start-up Time: 30 seconds from power-on
- Test Time: 60 seconds
- Dimensions: 35.5 cm x 35.5 cm x 38.1 cm