Description
A fully automated, multi-sample osmometer that sets the new benchmark for analytical performance, ease of use, and true walkaway operation
Specifications:
- Sample Test Volume: 100 µL
- Aspirated Sample Volume: 150 µL
- Sample Capacity: 20-sample carousel
- Test Time: Less than 3 minutes
- Sample Throughput: Can process 20 samples in less than 1 hour
- Units: mOsm/kg H2O
- Calibrated Range: Low: 0 to 2000 mOsm/kg H2O; Full: 0 to 4000 mOsm/kg H2O
- Resolution: 1 mOsm/kg H2O
- Calibration: 3-point calibration for low range, 4-point for full range
- Communications: Onboard printer, 10/100Mbps Ethernet, 4 USB 1.0/1.1/2.0, integrated bar code scanner
- Accuracy*: 0 to 400 mOsm: mean value ≤2 mOsm/ kg H2O from nominal value 400 to 4000 mOsm: mean value ≤0.5% from nominal value
- Precision*(within run): Standard deviation ±2 mOsm/kg H2O between 0 and 400 mOsm; standard deviation ± 0.5% of value between 400 and 4000 mOsm
- Drift*: Less than 1 mOsm/kg H2O per month
- Temperature Effect**: Recalibration recommended for ambient temperature changes of 5°C (9°F) or greater
- Storage Temperature: -40°C to +45°C (-40°F to +113°F)
- Sample Viscosity: Up to 20 mPa•s (20 cP), higher viscosities may affect pipettor accuracy
- Pipettor Accuracy: <1%
- Electrical Voltage: 100-240 V AC (50-60 Hz)
- Power Consumption: 375 W
- Dimensions(D x W x H): 20.5” x 23.6” x 22.8” , (52 cm x 60 cm x 58 cm)
- Net Weight: 68 lb (31 kg)
* Performance at Reference Conditions — 20°C to 25°C (68°F to 77°F); 40% to 60% relative humidity; tolerances of reference or calibration solutions excluded
** Operating Conditions — 18°C to 35°C (64°F to 95°F); 5% to 80% relative humidity (noncondensing)