Description
The 1100 HPLC G1321A Fluorescence Detector brings high-sensitivity and selection through simultaneous multi-wavelength detection. This easy-to-use detector provides quantitative data and fluorescence spectra from a single run. The FD’s spectral information can be used for rapid method optimization and verification of separation quality.
Specifications:
- Dimensions: 5 x 13.5 x 17 in
- Weight:5 lbs
- Environment:0 to 40°C at <95% humidity
- Performance:10 fg Anthracene, Ex=250 nm, Em=400 nm*
- Performance:RAMAN (H2O) > 200 (FLF rev A)
- Performance:RAMAN (H2O) > 400 (FLF rev >A)
- Performance:Ex=350 nm, Em=397 nm, dark value 450 nm, standard flow cell time constant=4 seconds (8 seconds response time)
- Pulse Frequency:296Hz (single signal mode), 74Hz (spectral mode)
- Excitation Monochromator Range:200 nm – 700 nm
- Excitation Monochromator Bandwidth:20 nm
- Emission Monochromator Range:280 nm – 900 nm
- Emission Monochromator Bandwidth:20 nm
- Spectrum Acquisition Scan Speed:28 ms/datapoint
- Wavelength Characteristic Repeatability:+/- 0.2 nm
- Wavelength Characteristic Accuracy: +/- 3 nm
- Flow cells: 8 uL (standard) and 20 bar (2MPa) pressure maximum