Agilent 1100 Series – G1377A HPLC Well Plate Autosampler

Cambridge Scientific ID: 13766 Category: Tag:

Description

Specifications:

  • GLP features: Early maintenance feedback (EMF), electronic records of maintenance and errors
  • Communications: Communications Controller-area network (CAN). RS232C, APG-remote standard, optional four external contact closures and BCD vial number output
  • Safety features: Leak detection and safe leak handling, low voltages in maintenance areas, error detection and display
  • Injection range:
    • – 8 µl in 0.01 µl increments with the small loop capillary
    • – 40 µl in 0.01 µl increments with the extended loop capillary
  • Precision:
    • Typically < 0.5 % RSD of peak areas from 5 – 40 µl,
    • Typically < 1 % RSD from 1 – 5 µl
    • Typically < 3 % RSD from 0.2 – 1 µl
  • Sample viscosity range: 2 – 5 cp
  • Sample capacity:
    • 2× well-plates (MTP) + 10 × 2 ml vials
    • 100 x 2 ml in one tray
    • 40 x 2 ml in half tray
  • Injection cycle time:
    • Typically, < 30 s using following standard conditions:
    • Default draw speed: 4 µl/min
    • Default eject speed: 10 µl/min
    • Injection volume: 0.1 µl
  • Carry-over:
    • Typically < 0.05 % using the following conditions:
    • Column: 150 x 0.5 mm Hypersil ODS, 3 µm
    • Mobile phase: Water/Acetonitrile = 85/15
    • Column Flow rate: 13 µl/min
    • Injection volume: 1 µl caffeine (=25ng caffeine), 1 µl water to test carryover
    • Outside wash of needle before injection: 20 sec with water using flush port

Contact Us

Thank you for your interest in the Agilent 1100 Series – G1377A HPLC Well Plate Autosampler (SKU:13766) To speak with a representative, please fill out the contact form below or call us directly.

Product Inquiry