Specifications:
- Detection type: 1024-element photodiode array
- Light source: Deuterium and tungsten lamps
- Wavelength range: 190 – 950 nm
- Short term noise (ASTM) Single and Multi-Wavelength: ± 1 × 10-5AU at 254 and 750 nm
- Drift: 2 × 10-3AU/hr at 254 nm
- Linear absorbance Range: 2 AU (upper limit)
- Wavelength accuracy: ± 1 nm
- Wavelength bunching: 1 – 400 nm
- Slit width: 1, 2, 4 , 8, 16 nm
- Diode width: < 1 nm
- Flow cells
- Standard: 13 µl volume, 10 mm cell path length and 120 bar (1760 psi) pressure maximum
- Semi-Micro: 5 µl volume, 6 mm cell path length and 120 bar (1760 psi) pressure maximum
- High pressure: 1.7 µl volume, 6 mm cellpath length and 400 bar (5880 psi) pressure maximum
- 500 nano: 0.5 µl volume, 10 mm cell path length and 50 bar (725 psi) pressure maximum
- Control and data Evaluation: Agilent ChemStation for LC
- Analog outputs: Recorder/integrator: 100 mV or 1 V, output range 0.001 – 2 AU, two outputs
- Communications: Controller-area network (CAN), GPIB, RS-232C, APG Remote: ready, start, stop and shut-down signals, LAN optional
- Safety and maintenance: Extensive diagnostics, error detection and display (through control module and ChemStation), leak detection, safe leak handling, leak output signal for shutdown of pumping system. Low voltages in major maintenance areas.
- GLP features: Early maintenance feedback (EMF) for continuous tracking of instrument usage in terms of lamp burn time with user-settable limits and feedback messages. Electronic records of maintenance and errors. Verification of wavelength accuracy with built-in holmium oxide filter
Can be sold as an individual component or as part of a complete stack