Description
Specifications:
- Column capacity:
- CM-A: Two columns, as standard (maximum length of 150 mm with filter or guard column)
- or four columns (maximum length of 50 mm) can be supported with optional tubing kit,
- up to 4.6 mm internal diameter (1.0.)
- CM-Aux: Two columns (maximum length of 150 mm, with filter or guard column) – up to
- two CM-Aux units can be configured with one CM-A for support of up to six columns
- Switching valves: Two nine-port, eight-position valves (CM-A only); provides programmable, automatic, random access switching, waste and bypass positions for rapid solvent changeover
- Temperature Range:
- 0 to 90.0 ·C, set table in 0.1 ·C increments
- Two independent heatl cool zones per module, up to six zones in stacked configuration
- Temperature Accuracy: ±0.5 ·C
- Temperature Stability: ±0.3 ·C
- Solvent conditioning: Active pre-heating as standard
- Column tracking: eCord Technology column information management tracks and archives column usage history
- 20 support: Optional