Description
Genevac HT Series 3i Vacuum Evaporator
Looking for a high-throughput vacuum evaporator that saves time and space? The Genevac HT Series 3i does all that and more! The ideal solution for parallel evaporation bottlenecks, our innovative high-performance Genevac HT Series 3i vacuum evaporators are suitable for high sample capacities, increasing your productivity.
- Dri-Pure anti bumping technology for foam-free, contamination-free evaporation
- Easy to use thanks to intuitive touchscreen controls
- Fast lyophilization Crystallization capability,
- ampleGuard temperature control protects samples from overheating to give peace of mind
- Sample flexibility – holders for tubes, vials, flasks, and 96 well plates and Samplegenie
Features & Benefits:
- Intuitive touchscreen controls: ‘Press & Go’ feature with simplified manual and automatic programming means easy set-up and run
- Save valuable fume hood space; integrated condenser means a smaller overall system footprint
- Front opening door allows for easy and error-free access
- Options for hydrochloric acid (HCL) resistance, Inert Gas purge (IGP), and EXALT controlled crystallization
- A wide range of sample holders for tubes, vials, flasks up to 500 mL, and bespoke FastStack holders for microtiter plates
- Minimized environmental impact thanks to a high-performance dry vacuum pump, and F-Gas compliant -75°C auto-defrost and draining condenser with R449A and R170 low Global Warming potential (GWP) refrigerants.
- Pre-programmed methods for ease of use, with user editable parameters enabling users to fine tune your methods for optimal performance
Specifications:
Evaporator:
- Max. rotor speed: 1415 rpm
- Max. load per swing: 1.5 kg
- Max. operational imbalance: 80 g
- Dimensions (W x D x H): 660 x 710 x 840 mm
- Weight: 193.3 kg
Vacuum Pump (External):
- Type: Oil-free Scroll
- Ultimate system vacuum: <0.4 mbar
- Dimensions (W xD x H): 432 x 282 x 302 mm
- Weight: 36.2 kg
- Vacuum hose/control cable: 3 m
Condenser:
- Type: Dual-stage vapour compression
- Refrigerant gas – stage 1: R449A
- Refrigerant charge – stage 1: 320 g
- Refrigerant GWP – stage 1: 1,397
- Refrigerant CO2e – stage 1: 0.5 tonnes
- Refrigerant gas – stage 2: R170
- Refrigerant charge – stage 2: 41g
- Refrigerant GWP – stage 2: 6
- Total CO2e – stage 2: < 0.001 tonnes
- Ultimate low temperature²: -75°C
- Max. Pressure (PS): 30 bar
Emissions:
- Noise (@ 1 metre): 65 dB(A)
- Exhaust hose(supplied): 6 mm ID / 8 mm OD
Electrical:
- Supply: 230 V 50 Hz, 220 V 60 Hz, 208 V 60 Hz
- Max supply input: 1500 A
Power Consumption: Current (A) at unit voltage
Operational Environment:
- Ambient temperature: 15°C to 30°C
- Relative humidity: 10-80% non-condensing
- Altitude: Sea-level to 1600 m
- Min. ventilation air-gap: 70 mm
- Installation environment: Indoor only. Static-dissipative laboratory or similar
Solvent Capacity & ACC Range:
- Max. solvent capacity: 4.5 L
- Refrigeration ACC range: 100°C
Inter Gas Supply Requirements:
- Max. pressure: 2 bar g
- Min. pressure: 1.5 bar g
- Flow rate: 50 litres/min
- Hose length: 2.5 m
- Max. consumption(purge): 120 litres
- Max. consumption(blanket): 60 litres/hour
- Connector type: ⅜ “ BSP female