Cambridge Scientific
Used Lab Equipment

Used Lab Equipment

We offer an assortment of refurbished and used Liquid Chromatography solutions ranging from HPLC, LC/MS to FPLC. Our systems can be customized to include almost any configuration of Autosamplers, Detectors and Pumps. Some of our leading manufacturers include Agilent, HP, Waters, GE, Amersham, Dionex and Shimazdzu. Our HPLC systems such as the Agilent 1100 and 1200 Series, Akta FPLCs, and the Waters Alliance & 600 series have the option of on site training, installation and extended warranties.

Hplc Variable Wavelength Detector: 1-5 of 5

*Please Select a Category and/or Manufacturer dropdown OR Search By Model Name, # or Keyword.* Advanced Search >
Agilent 1200 Series - G1314B HPLC Variable Wavelength DetectorImage Coming Soon

Agilent 1200 Series - G1314B HPLC Variable Wavelength Detector

Cambridge ID #: 18527

The Agilent 1200 Series G1314B variable wavelength detector is designed for highest optical performance, GLP compliance and easy maintenance with higher data rate (27/55Hz) for fast-HPLC, deuterium lamp for highest intensity and lowest detection limit over a wavelength range of 190 to 600 nm, the detector also features easy front access to lamp and flow cell for fast replacement, and built-in holmium oxide filter for fast wavelength accuracy verification.

 

Specification:

 

  • Detection type: Double-beam photometer
  • Light source: Deuterium lamp
  • Wavelength range: 190-600 nm
  • Short term noise (ASTM): ± 0.75 x 10^-5 AU at 254 nm
  • Drift: 3 x 10^-4 AU/hr at 254 nm
  • Linearity: > 2 AU (5%) upper limit
  • Wavelength accuracy: ± 1 nm
  • Band with: 6.5 nm typical

 

  • Flow cells

 

    • Standard: 14-µl volume, 10-mm cell path length and 40 bar (588 psi) pressure maximum
    • High pressure: 14-µl volume, 10-mm cell path length and 400 bar (5880 psi) pressure maximum
    • Micro: 1-µl volume, 5-mm cell path length and 40 bar (588 psi) pressure maximum
    • Semi-micro: 5-µl volume, 6-mm cell path length and 40 bar (588 psi) pressure maximum
  • Control and data evaluation: Agilent ChemStation for LC
  • Analog outputs: Recorder/integrator: 100 mV or 1 V, output range 0.001 – 2 AU, one output
  • Communications: Controller-area network (CAN), GPIB, RS-232C, APG Remote: ready, start, stop and shut-down signals, LAN optional
  • Safety and maintenance: Extensive diagnostics, error detection and display (through control module and Agilent ChemStation), leak detection, safe leak handling, leak output signal for shutdown of pumping system. Low voltages in major maintenance areas.
  • GLP features: Early maintenance feedback (EMF) for continuous tracking of instrument usage in terms of lamp burn time with user-settable limits and feedback messages. Electronic records of maintenance and errors. Verification of wavelength accuracy with built-in holmium oxide filter.
  • Housing: All materials recyclable
Learn More
Agilent-1200-Series---G1314B-HPLC-Variable-Wavelength-Detector-18527
By clicking below we can further assist you with your inquiry. Inquire 1 In Stock Worldwide Shipping
Dionex Ultimate 3000 VWD-3100 HPLC Variable Wavelength DetectorImage Coming Soon

Dionex Ultimate 3000 VWD-3100 HPLC Variable Wavelength Detector

Cambridge ID #: 17926

Specifications:

  • Optical design: Dual-beam photometer
  • Light source: Deuterium lamp and tungsten lamp
  • Wavelength range: 190 to 900 nm
  • Wavelength accuracy: ± 1 nm
  • Wavelength repeatability: y: ± 0.1 nm
  • Optical bandwidth: 6 nm at 254 nm
Learn More
Dionex-Ultimate-3000-VWD-3100-HPLC-Variable-Wavelength-Detector-17926
By clicking below we can further assist you with your inquiry. Inquire 1 In Stock Worldwide Shipping
Agilent 1100 Series - G1314A HPLC Variable Wavelength Detector

Agilent 1100 Series - G1314A HPLC Variable Wavelength Detector

Cambridge Scientific ID #'s: 16036, 10251, 10893, 18127

Specification:

  • Detection type: Double-beam photometer
  • Light source: Deuterium lamp
  • Wavelength range: 190–600 nm
  • Short term noise (ASTM): ±0.75×10-5 AU at 254 nm
  • Drift: 3 × 10 -4 AU/hr at 254 nm
  • Linearity: 2 AU (5%) upper limit
  • Wavelength accuracy:
  • Band width: 5 nm typical
  • Flow cells:
    • Standard: 14-μl volume, 10-mm cell path length and 40 bar (588 psi) pressure maximum
    • High pressure: 14-μl volume, 10-mm cell path length and 400 bar (5880 psi) pressure maximum
    • Micro: 1-μl volume, 5-mm cell path length and 40 bar (588 psi) pressure maximum
    • Semimicro: 5-μl volume, 6-mm cell path length and 40 bar (588 psi) pressure maximum
  • Control and data evaluation: Agilent ChemStation for LC
  • Analog outputs: Recorder/integrator: 100 mV or 1 V, output range 0.001 – 2 AU, one output
  • Communications: Controller-area network (CAN), GPIB, RS-232C, APG Remote: ready, start, stop and shut-down signals, LAN optional
  • Safety and maintenance: Extensive diagnostics, error detection and display (through control module and Agilent ChemStation), leak detection, safe leak handling, leak output signal for shutdown of pumping system. Low voltages in major maintenance areas.
  • GLP features: Early maintenance feedback (EMF) for continuous tracking of instrument usage in terms of lamp burn time with user-settable limits and feedback messages. Electronic records of maintenance and errors. Verification of wavelength accuracy with built-in holmium oxide filter
  • Housing: All materials recyclable.
Learn More
Agilent-1100-Series---G1314A-HPLC-Variable-Wavelength-Detector-16036
By clicking below we can further assist you with your inquiry. Inquire 4 In Stock Worldwide Shipping
Agilent 1200 Series - G1314C HPLC Variable Wavelength Detector

Agilent 1200 Series - G1314C HPLC Variable Wavelength Detector

Cambridge ID #: 15899

Specifications: 

  • Detection type: Double-beam photometer
  • Light source: Deuterium lamp
  • Noise: ± 0.75 x 10 -5 AU, at 254 nm
  • Drift: 3 x 10-4AU/hr, at 254 nm
  • Linearity: >2 AU upper limit
  • Wavelength range: 190-600 nm
  • Wavelength accuracy: ± 1 nm, self-calibration with deuterium lines, verification with holmium oxide filter
  • Wavelength bandwidth: 6.5 nm typical over whole wavelength range
  • Maximum data rate: 55 Hz
Learn More
Agilent-1200-Series---G1314C-HPLC-Variable-Wavelength-Detector-15899
By clicking below we can further assist you with your inquiry. Inquire 1 In Stock Worldwide Shipping
Dionex Ultimate 3000 VWD-3400RS HPLC Variable Wavelength Detector

Dionex Ultimate 3000 VWD-3400RS HPLC Variable Wavelength Detector

Cambridge ID #: 15269

Specifications:

  • Optical design: Dual-beam photometer
  • Light source: Deuterium lamp and tungsten lamp
  • Wavelength range: 190 to 900 nm
  • Wavelength accuracy: ± 1 nm
  • Wavelength repeatability: y: ± 0.1 nm
  • Optical bandwidth: 6 nm at 254 nm
Learn More
Dionex-Ultimate-3000-VWD-3400RS-HPLC-Variable-Wavelength-Detector-15269
By clicking below we can further assist you with your inquiry. Inquire 1 In Stock Worldwide Shipping